Recall Z-0816-2024— American Contract Systems, Inc.
Class II · Ongoing
Class II FDA medical device recall by American Contract Systems, Inc., initiated 2023-11-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Custom procedural convenience kits and trays, anesthesia, labeled as: a) HEART ANESTHESIA BUNDLE NS, Kit Number LMHA01V; b) OR/Anesthesia Central Line Insertion Kit, Kit Number UIAN43L. Reason: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals..
- Recalling firm
- American Contract Systems, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-11-29
- Firm location
- Temple Terrace, FL, United States
Product description
Custom procedural convenience kits and trays, anesthesia, labeled as: a) HEART ANESTHESIA BUNDLE NS, Kit Number LMHA01V; b) OR/Anesthesia Central Line Insertion Kit, Kit Number UIAN43L
Reason for recall
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Data sourced from openFDA. This site is unofficial and independent of the FDA.