Recall Z-0813-2019— Stryker Medical Division of Stryker Corporation

Class II · Terminated

Class II FDA medical device recall by Stryker Medical Division of Stryker Corporation, initiated 2018-10-10, terminated 2020-04-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Naera Hospital Bassinet, Model No. 4402. Reason: It was identified that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub. A Health Hazard Evaluation was completed which identified the potential risk of patients moving out of position and rolling over on the mattress. While the highest possible severity of harm associated with this condition is death, there have been no reports of injury or harm. This condition has only presented in regions of high altitude; however, Stryker has chosen to replace all Naera mattresses in the field..

Recalling firm
Stryker Medical Division of Stryker Corporation
Classification
Class II
Status
Terminated
Initiated
2018-10-10
Terminated
2020-04-29
Firm location
Portage, MI, United States

Product description

Naera Hospital Bassinet, Model No. 4402

Reason for recall

It was identified that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub. A Health Hazard Evaluation was completed which identified the potential risk of patients moving out of position and rolling over on the mattress. While the highest possible severity of harm associated with this condition is death, there have been no reports of injury or harm. This condition has only presented in regions of high altitude; however, Stryker has chosen to replace all Naera mattresses in the field.

Data sourced from openFDA. This site is unofficial and independent of the FDA.