Recall Z-0812-2024— Philips North America
Class III · Ongoing
Class III FDA medical device recall by Philips North America, initiated 2023-12-29. It names one FDA 510(k) clearance: K193215. Product: SmartPath to dStream for 1.5T, Model No. 781260, 782112. Reason: An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel..
- Recalling firm
- Philips North America
- Classification
- Class III
- Status
- Ongoing
- Initiated
- 2023-12-29
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
SmartPath to dStream for 1.5T, Model No. 781260, 782112
Reason for recall
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.