Recall Z-0811-2024— Philips North America

Class III · Ongoing

Class III FDA medical device recall by Philips North America, initiated 2023-12-29. It names one FDA 510(k) clearance: K193215. Product: Intera 1.5T, Model No. 781195. Reason: An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel..

Recalling firm
Philips North America
Classification
Class III
Status
Ongoing
Initiated
2023-12-29
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Intera 1.5T, Model No. 781195

Reason for recall

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.