Recall Z-0810-2018— Howmedica Osteonics Corp.
Class II · Terminated
Class II FDA medical device recall by Howmedica Osteonics Corp., initiated 2017-09-08, terminated 2018-07-31. FDA's recall record names no specific 510(k) clearance for this event. Product: AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to the implant and is specifically designed to allow a surgeon to insert the implant into the disc space with a fixed engagement between the two components to permit a controlled positioning of the implant in the anterior portion of the intervertebral space, and then through activation of the rotational pivot between the Handle and implant for final placement of the crescent-shaped cage across the anterior portion of the disc space. The AccuLIF TL Cages are indicated for intervertebral body fusion with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DOD) at one level or two contiguous levels from L2 to S1.. Reason: When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection..
- Recalling firm
- Howmedica Osteonics Corp.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-09-08
- Terminated
- 2018-07-31
- Firm location
- Allendale, NJ, United States
Product description
AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to the implant and is specifically designed to allow a surgeon to insert the implant into the disc space with a fixed engagement between the two components to permit a controlled positioning of the implant in the anterior portion of the intervertebral space, and then through activation of the rotational pivot between the Handle and implant for final placement of the crescent-shaped cage across the anterior portion of the disc space. The AccuLIF TL Cages are indicated for intervertebral body fusion with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DOD) at one level or two contiguous levels from L2 to S1.
Reason for recall
When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection.
Data sourced from openFDA. This site is unofficial and independent of the FDA.