Recall Z-0810-2013— Mizuho OSI

Class II · Terminated

Class II FDA medical device recall by Mizuho OSI, initiated 2012-12-17, terminated 2016-04-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.. Reason: Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top..

Recalling firm
Mizuho OSI
Classification
Class II
Status
Terminated
Initiated
2012-12-17
Terminated
2016-04-15
Firm location
Union City, CA, United States

Product description

Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.

Reason for recall

Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.

Data sourced from openFDA. This site is unofficial and independent of the FDA.