Recall Z-0809-2022— Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Biomet, Inc., initiated 2022-02-14, terminated 2024-09-30. It names one FDA 510(k) clearance: K141263. Product: BIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.. Reason: Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part..

Recalling firm
Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2022-02-14
Terminated
2024-09-30
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.

Reason for recall

Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.