Recall Z-0807-2020— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2019-12-20, terminated 2020-05-01. It names one FDA 510(k) clearance: K163286. Product: Siemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: Angiography system for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.. Reason: The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred..

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2019-12-20
Terminated
2020-05-01
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Siemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: Angiography system for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason for recall

The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred.

Data sourced from openFDA. This site is unofficial and independent of the FDA.