Recall Z-0805-2015— American Medical Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by American Medical Systems, Inc., initiated 2014-10-15, terminated 2015-05-28. It names 2 FDA 510(k) clearances: K131229, K051530. Product: AMS Monarc + Subfascial Hammock with Tensioning Suture REF 72404193 and AMS Monarc C Subfascial Hammock with Tensioning Suture REF 72404195, Sterilized Using Ethylene Oxide. The Monarc + and Monarc C subfascial hammocks are sterile, single use procedure kits, each consisting of two stainless steel, curved needle passers. The tip portion of each needle passer is configured to allow for secure placement of the connectors. Each needle passer has a plastic handle attached. The system also contains a sling assembly including one piece of loosely knitted polypropylene mesh, two removable plastic insertion sheaths attached to the mesh, and two connectors attached to the insertion sheaths. The mesh is constructed of polypropylene monofilament. An absorbable tensioning suture is threaded into the length of the mesh to allow for tensioning adjustment of the mesh after placement in the patient is achieved. The plastic covering also affords convenient travel of the mesh through the tissue. The connectors are attached to the tip portions of the Monarc needle passers during the procedure. The loosely knitted polypropylene mesh is intended to remain in the body as a permanent implant. The mesh component is not absorbed or degraded by the action of tissue in-growth or tissue enzymes.. Reason: During routine periodic packaging testing, AMS identified that in simulated extreme shipping and handling situations the sterile barrier packaging of the Needle Passer components supplied with the AdVance" Male Sling System; Model # 720088-01; AdVance" XP Male Sling System; Model # 720163-01; Monarc" C Subfacial Hammock; Model # 72404195; Monarc" + Subfacial Hammock; Model #.

Recalling firm
American Medical Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-10-15
Terminated
2015-05-28
Firm location
Minnetonka, MN, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AMS Monarc + Subfascial Hammock with Tensioning Suture REF 72404193 and AMS Monarc C Subfascial Hammock with Tensioning Suture REF 72404195, Sterilized Using Ethylene Oxide. The Monarc + and Monarc C subfascial hammocks are sterile, single use procedure kits, each consisting of two stainless steel, curved needle passers. The tip portion of each needle passer is configured to allow for secure placement of the connectors. Each needle passer has a plastic handle attached. The system also contains a sling assembly including one piece of loosely knitted polypropylene mesh, two removable plastic insertion sheaths attached to the mesh, and two connectors attached to the insertion sheaths. The mesh is constructed of polypropylene monofilament. An absorbable tensioning suture is threaded into the length of the mesh to allow for tensioning adjustment of the mesh after placement in the patient is achieved. The plastic covering also affords convenient travel of the mesh through the tissue. The connectors are attached to the tip portions of the Monarc needle passers during the procedure. The loosely knitted polypropylene mesh is intended to remain in the body as a permanent implant. The mesh component is not absorbed or degraded by the action of tissue in-growth or tissue enzymes.

Reason for recall

During routine periodic packaging testing, AMS identified that in simulated extreme shipping and handling situations the sterile barrier packaging of the Needle Passer components supplied with the AdVance" Male Sling System; Model # 720088-01; AdVance" XP Male Sling System; Model # 720163-01; Monarc" C Subfacial Hammock; Model # 72404195; Monarc" + Subfacial Hammock; Model #