Recall Z-0801-2019— Radiometer America Inc
Class II · Terminated
Class II FDA medical device recall by Radiometer America Inc, initiated 2018-07-02, terminated 2024-04-02. It names one FDA 510(k) clearance: K051968. Product: Product: ABL800 FLEX with Crea. The ABL800 with Crea is available in the following configurations: ABL817, ABL827 and ABL837. ABL800 Series Ref. 393-803, 393-804, 393-805, 393-806, 393-820, 393-821, 393-825, 393-826, 393-835, 393-836, 393-844, 393-845.. Reason: From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine..
- Recalling firm
- Radiometer America Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-07-02
- Terminated
- 2024-04-02
- Firm location
- Brea, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Product: ABL800 FLEX with Crea. The ABL800 with Crea is available in the following configurations: ABL817, ABL827 and ABL837. ABL800 Series Ref. 393-803, 393-804, 393-805, 393-806, 393-820, 393-821, 393-825, 393-826, 393-835, 393-836, 393-844, 393-845.
Reason for recall
From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine.