Recall Z-0794-2020— Baxter Healthcare Corporation

Class II · Ongoing

Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2019-12-12. It names one FDA 510(k) clearance: K961225. Product: ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8377, 7N8378, 7N8390, 7N8391, 7N8399 - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use. This device is an in-line injection site which can be connected to standard male Luer adapters (e.g., syringes or sets) for the continuous or intermittent fluid administration or the withdrawal of fluids.. Reason: Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector..

Recalling firm
Baxter Healthcare Corporation
Classification
Class II
Status
Ongoing
Initiated
2019-12-12
Firm location
Deerfield, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8377, 7N8378, 7N8390, 7N8391, 7N8399 - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use. This device is an in-line injection site which can be connected to standard male Luer adapters (e.g., syringes or sets) for the continuous or intermittent fluid administration or the withdrawal of fluids.

Reason for recall

Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.