Recall Z-0791-2014— Philips Medical Systems (Cleveland) Inc
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2011-03-07, terminated 2014-04-28. It names 3 FDA 510(k) clearances: K060937, K033357, K033326. Product: Brilliance iCT, Brilliance 64 and Brilliance Big Bore The Brilliance iCT, Brilliance 64 and Brilliance Big Bore are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.. Reason: The DoseRight feature suggest a mAs based on the measured patient size, a reference size and a reference mAs. When scanning large children, the suggested mAs may be higher than clinicians would expect..
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2011-03-07
- Terminated
- 2014-04-28
- Firm location
- Cleveland, OH, United States
510(k) clearances named in this recall
3 on FDA's record- K060937 — BRILLIANCE VOLUME
- K033357 — ACQSIM-MULTISLICE-CT, MODEL 453567042271
- K033326 — PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brilliance iCT, Brilliance 64 and Brilliance Big Bore The Brilliance iCT, Brilliance 64 and Brilliance Big Bore are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Reason for recall
The DoseRight feature suggest a mAs based on the measured patient size, a reference size and a reference mAs. When scanning large children, the suggested mAs may be higher than clinicians would expect.