Recall Z-0783-2014— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2013-12-19, terminated 2015-02-24. It names one FDA 510(k) clearance: K070978. Product: Zimmer Segmental Systems; Polyethylene Inserts, Size B. Designed to address significant bone loss resulting from oncology, trauma, and/or salvage of previously failed arthroplasty.. Reason: After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-12-19
Terminated
2015-02-24
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Zimmer Segmental Systems; Polyethylene Inserts, Size B. Designed to address significant bone loss resulting from oncology, trauma, and/or salvage of previously failed arthroplasty.

Reason for recall

After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.