Recall Z-0774-2013— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2012-09-14, terminated 2014-06-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Trilogy Acetabular AB Shells with Cluster Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.. Reason: Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-09-14
Terminated
2014-06-20
Firm location
Warsaw, IN, United States

Product description

Trilogy Acetabular AB Shells with Cluster Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.

Reason for recall

Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.

Data sourced from openFDA. This site is unofficial and independent of the FDA.