Recall Z-0771-2016— Accuray Incorporated

Class II · Terminated

Class II FDA medical device recall by Accuray Incorporated, initiated 2016-01-22, terminated 2016-10-20. It names one FDA 510(k) clearance: K120233. Product: CyberKnife Robotic; Catalog/part number 032000 and 033000 Cosmetic cover package. Product Usage: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.. Reason: The gun box mounting bracket may fail to support the weight of the gun box when in vertical (inverted) position. If this failure occurs the gun box may become loose and could come into contact with a patient..

Recalling firm
Accuray Incorporated
Classification
Class II
Status
Terminated
Initiated
2016-01-22
Terminated
2016-10-20
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CyberKnife Robotic; Catalog/part number 032000 and 033000 Cosmetic cover package. Product Usage: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Reason for recall

The gun box mounting bracket may fail to support the weight of the gun box when in vertical (inverted) position. If this failure occurs the gun box may become loose and could come into contact with a patient.

Data sourced from openFDA. This site is unofficial and independent of the FDA.