Recall Z-0769-2017— Natus Medical Incorporated
Class II · Terminated
Class II FDA medical device recall by Natus Medical Incorporated, initiated 2016-09-16, terminated 2017-09-01. It names one FDA 510(k) clearance: K103589. Product: Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS. Reason: The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box..
- Recalling firm
- Natus Medical Incorporated
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-09-16
- Terminated
- 2017-09-01
- Firm location
- Seattle, WA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
Reason for recall
The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.
Data sourced from openFDA. This site is unofficial and independent of the FDA.