Recall Z-0765-2025— MEDLINE INDUSTRIES, LP - Northfield

Class I · Ongoing

Class I FDA medical device recall by MEDLINE INDUSTRIES, LP - Northfield, initiated 2024-12-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;. Reason: Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber..

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Classification
Class I
Status
Ongoing
Initiated
2024-12-16
Firm location
Northfield, IL, United States

Product description

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;

Reason for recall

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Data sourced from openFDA. This site is unofficial and independent of the FDA.