Recall Z-0764-2025— MEDLINE INDUSTRIES, LP - Northfield

Class I · Ongoing

Class I FDA medical device recall by MEDLINE INDUSTRIES, LP - Northfield, initiated 2024-12-16. It names one FDA 510(k) clearance: K852140. Product: Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223. Reason: Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber..

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Classification
Class I
Status
Ongoing
Initiated
2024-12-16
Firm location
Northfield, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223

Reason for recall

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Data sourced from openFDA. This site is unofficial and independent of the FDA.