Recall Z-0763-2019— Edwards Lifesciences, LLC

Class I · Terminated

Class I FDA medical device recall by Edwards Lifesciences, LLC, initiated 2018-12-21, terminated 2023-11-02. It names one FDA 510(k) clearance: K160084. Product: Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.. Reason: These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms..

Recalling firm
Edwards Lifesciences, LLC
Classification
Class I
Status
Terminated
Initiated
2018-12-21
Terminated
2023-11-02
Firm location
Irvine, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

Reason for recall

These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.

Data sourced from openFDA. This site is unofficial and independent of the FDA.