Recall Z-0762-2025— Moximed, Inc.

Class II · Ongoing

Class II FDA medical device recall by Moximed, Inc., initiated 2024-12-02. FDA's recall record names no specific 510(k) clearance for this event. Product: MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.. Reason: Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device..

Recalling firm
Moximed, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-12-02
Firm location
Fremont, CA, United States

Product description

MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.

Reason for recall

Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.

Data sourced from openFDA. This site is unofficial and independent of the FDA.