Recall Z-0762-2020— Deerfield Imaging, Inc.
Class II · Terminated
Class II FDA medical device recall by Deerfield Imaging, Inc., initiated 2019-11-26, terminated 2020-08-04. It names one FDA 510(k) clearance: K101813. Product: iMRX MRI and Angiography System, part number 10094137 and 10094141, a surgical/interventional suite with integrated MRI and X-ray angiography. The IMRIS iMRX system is the integration of a Siemens Artis zee family (biplane or floor or ceiling mounted single plane) and the IMRIS Neuro 3T System. The System is a solution providing MRI scanning capabilities in the angiography room while retaining all functionality of Siemens Artis zee system. The System includes a traditional MRI unit that has been suspended on an overhead rail system, and is designed to operate inside RF room to facilitate intra-operative, interventional, and multi-room imaging. It also allows intra-operative or interventional imaging capabilities, while retaining all diagnostic functionality of a standard MRI system in the diagnostic room. The iMRX system allows for the sequential use of either MR imaging or X-ray fluoroscopic imaging of the head or chest for a single patient while positioned on a common fixed table either prior to, during or post clinical procedure. When these images are interpreted by a trained physician, they may assist in diagnosis during either neurovascular or cardiovascular interventionsa procedures. The iMRX system is designed such that the imaging sub-systems cannot operate simultaneously and only in accordance with their respective cleared intended use.. Reason: There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible.
- Recalling firm
- Deerfield Imaging, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-11-26
- Terminated
- 2020-08-04
- Firm location
- Minnetonka, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
iMRX MRI and Angiography System, part number 10094137 and 10094141, a surgical/interventional suite with integrated MRI and X-ray angiography. The IMRIS iMRX system is the integration of a Siemens Artis zee family (biplane or floor or ceiling mounted single plane) and the IMRIS Neuro 3T System. The System is a solution providing MRI scanning capabilities in the angiography room while retaining all functionality of Siemens Artis zee system. The System includes a traditional MRI unit that has been suspended on an overhead rail system, and is designed to operate inside RF room to facilitate intra-operative, interventional, and multi-room imaging. It also allows intra-operative or interventional imaging capabilities, while retaining all diagnostic functionality of a standard MRI system in the diagnostic room. The iMRX system allows for the sequential use of either MR imaging or X-ray fluoroscopic imaging of the head or chest for a single patient while positioned on a common fixed table either prior to, during or post clinical procedure. When these images are interpreted by a trained physician, they may assist in diagnosis during either neurovascular or cardiovascular interventionsa procedures. The iMRX system is designed such that the imaging sub-systems cannot operate simultaneously and only in accordance with their respective cleared intended use.
Reason for recall
There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible