Recall Z-0762-2013— Roche Molecular Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Molecular Systems, Inc., initiated 2012-11-28, terminated 2015-08-06. FDA's recall record names no specific 510(k) clearance for this event. Product: COBAS TaqMan Mycobacteria tuberculosis (MTB) Test FOR IN VITRO DIAGNOSTIC USE.. Reason: The COBAS TaqMan MTB Test Definition Files (TDF) v3.0, for both CE-IVD and JPN-IVD tests, being used with AMPLILINK software version 3.3 series and COBAS TaqMan MTB Test Kits, both CE-IVD and JPN-IVD, have an incorrect Target Elbow Max cutoff value of 48 creating the potential for false negative results to be generated for patient samples with Target Elbows in the range of 48 to 50..

Recalling firm
Roche Molecular Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-11-28
Terminated
2015-08-06
Firm location
Branchburg, NJ, United States

Product description

COBAS TaqMan Mycobacteria tuberculosis (MTB) Test FOR IN VITRO DIAGNOSTIC USE.

Reason for recall

The COBAS TaqMan MTB Test Definition Files (TDF) v3.0, for both CE-IVD and JPN-IVD tests, being used with AMPLILINK software version 3.3 series and COBAS TaqMan MTB Test Kits, both CE-IVD and JPN-IVD, have an incorrect Target Elbow Max cutoff value of 48 creating the potential for false negative results to be generated for patient samples with Target Elbows in the range of 48 to 50.

Data sourced from openFDA. This site is unofficial and independent of the FDA.