Recall Z-0761-2017— Physio-Control, Inc.

Class II · Terminated

Class II FDA medical device recall by Physio-Control, Inc., initiated 2010-03-11, terminated 2016-12-16. It names one FDA 510(k) clearance: K082937. Product: The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. Product Part Numbers V-15-2-000XXX. Reason: Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010..

Recalling firm
Physio-Control, Inc.
Classification
Class II
Status
Terminated
Initiated
2010-03-11
Terminated
2016-12-16
Firm location
Redmond, WA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. Product Part Numbers V-15-2-000XXX

Reason for recall

Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010.