Recall Z-0758-2023— GE Medical Systems, LLC

Class II · Ongoing

Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2022-11-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative (abdominal, thoracic, and vascular).. Reason: Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire..

Recalling firm
GE Medical Systems, LLC
Classification
Class II
Status
Ongoing
Initiated
2022-11-10
Firm location
Waukesha, WI, United States

Product description

Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative (abdominal, thoracic, and vascular).

Reason for recall

Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.

Data sourced from openFDA. This site is unofficial and independent of the FDA.