Recall Z-0757-2014— The Anspach Effort, Inc.

Class II · Terminated

Class II FDA medical device recall by The Anspach Effort, Inc., initiated 2012-03-29, terminated 2015-03-11. It names one FDA 510(k) clearance: K042783. Product: MCA5-1SD - 1MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.. Reason: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide.

Recalling firm
The Anspach Effort, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-03-29
Terminated
2015-03-11
Firm location
Palm Beach Gardens, FL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

MCA5-1SD - 1MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Reason for recall

The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide

Data sourced from openFDA. This site is unofficial and independent of the FDA.