Recall Z-0756-2015— Stryker Endoscopy

Class II · Terminated

Class II FDA medical device recall by Stryker Endoscopy, initiated 2014-11-17, terminated 2015-07-17. FDA's recall record names no specific 510(k) clearance for this event. Product: PKG, ASSY, 5MM PEEK MULTI HANDLE, 45CM, P/N 0250181108 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.. Reason: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods.

Recalling firm
Stryker Endoscopy
Classification
Class II
Status
Terminated
Initiated
2014-11-17
Terminated
2015-07-17
Firm location
San Jose, CA, United States

Product description

PKG, ASSY, 5MM PEEK MULTI HANDLE, 45CM, P/N 0250181108 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason for recall

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Data sourced from openFDA. This site is unofficial and independent of the FDA.