Recall Z-0754-2013— GE Healthcare, LLC
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2012-05-24, terminated 2014-01-16. It names one FDA 510(k) clearance: K052345. Product: AdvantageSim MD versions 7.4 through 7.6, Model 5160092-2 Version 7.4, 5160092-3 Version 7.5, 5160092-4 version 7.6.. Reason: It was reported by a customer site that when using GE AdvantageSim MD on Advantage Workstation, the series could be incorrectly labeled in image view when multiple series of an exam are loaded simultaneously in Advantage Sim MD and if their series dates are different. A mismatch of series label for structure sets may lead to under-treatment of a tumor due to too small coverage of volume treated..
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-05-24
- Terminated
- 2014-01-16
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
AdvantageSim MD versions 7.4 through 7.6, Model 5160092-2 Version 7.4, 5160092-3 Version 7.5, 5160092-4 version 7.6.
Reason for recall
It was reported by a customer site that when using GE AdvantageSim MD on Advantage Workstation, the series could be incorrectly labeled in image view when multiple series of an exam are loaded simultaneously in Advantage Sim MD and if their series dates are different. A mismatch of series label for structure sets may lead to under-treatment of a tumor due to too small coverage of volume treated.