Recall Z-0753-2025— Abbott Medical

Class II · Ongoing

Class II FDA medical device recall by Abbott Medical, initiated 2024-11-21. FDA's recall record names no specific 510(k) clearance for this event. Product: The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.. Reason: Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing..

Recalling firm
Abbott Medical
Classification
Class II
Status
Ongoing
Initiated
2024-11-21
Firm location
Sylmar, CA, United States

Product description

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.

Reason for recall

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.

Data sourced from openFDA. This site is unofficial and independent of the FDA.