Recall Z-0753-2013— Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Biomet, Inc., initiated 2013-01-03, terminated 2013-08-02. It names one FDA 510(k) clearance: K934469. Product: TUNNELOC ROUND HEAD INTERFERENCE SCREW 8MMX25MM Biomet Sports Medicine manufactures a variety of internal fixation devices intended to aid in arthroscopic and orthopedic reconstructive procedures requiring soft tissue fixation, due to injury or degenerative disease. Implants used for this application include: screws, washers, anchors, pins, and suture. Interference Screws and Set Screws are intended for use in fixation of patellar bone-tendon-bone grafts in ACL reconstruction.. Reason: Thru hole at the tip of the screw is undersized..
- Recalling firm
- Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-01-03
- Terminated
- 2013-08-02
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
TUNNELOC ROUND HEAD INTERFERENCE SCREW 8MMX25MM Biomet Sports Medicine manufactures a variety of internal fixation devices intended to aid in arthroscopic and orthopedic reconstructive procedures requiring soft tissue fixation, due to injury or degenerative disease. Implants used for this application include: screws, washers, anchors, pins, and suture. Interference Screws and Set Screws are intended for use in fixation of patellar bone-tendon-bone grafts in ACL reconstruction.
Reason for recall
Thru hole at the tip of the screw is undersized.