Recall Z-0752-2024— Randox Laboratories Ltd.

Class II · Ongoing

Class II FDA medical device recall by Randox Laboratories Ltd., initiated 2023-11-20. It names one FDA 510(k) clearance: K053153. Product: Calibration Serum Level 3 CAL2351. Reason: CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%..

Recalling firm
Randox Laboratories Ltd.
Classification
Class II
Status
Ongoing
Initiated
2023-11-20
Firm location
Crumlin (North), N/A, Ireland

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Calibration Serum Level 3 CAL2351

Reason for recall

CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.