Recall Z-0752-2014— The Anspach Effort, Inc.
Class II · Terminated
Class II FDA medical device recall by The Anspach Effort, Inc., initiated 2012-03-29, terminated 2015-03-11. It names one FDA 510(k) clearance: K042783. Product: MCA5-07SD - 0.7MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.. Reason: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide.
- Recalling firm
- The Anspach Effort, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-03-29
- Terminated
- 2015-03-11
- Firm location
- Palm Beach Gardens, FL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
MCA5-07SD - 0.7MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.
Reason for recall
The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
Data sourced from openFDA. This site is unofficial and independent of the FDA.