Recall Z-0751-2024— Randox Laboratories Ltd.
Class II · Ongoing
Class II FDA medical device recall by Randox Laboratories Ltd., initiated 2023-11-13. It names one FDA 510(k) clearance: K053153. Product: Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532). Reason: There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human Assayed Multi-Sera Level 2 (HN1530) for TBIC taget value and on IFU of Human Assayed Multi-Sera Level 3 (HE 1532) for Lipase target value..
- Recalling firm
- Randox Laboratories Ltd.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-11-13
- Firm location
- Crumlin (North), N/A, Ireland
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532)
Reason for recall
There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human Assayed Multi-Sera Level 2 (HN1530) for TBIC taget value and on IFU of Human Assayed Multi-Sera Level 3 (HE 1532) for Lipase target value.