Recall Z-0750-2017— Roche Molecular Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Molecular Systems, Inc., initiated 2016-11-16, terminated 2017-12-12. FDA's recall record names no specific 510(k) clearance for this event. Product: LightMix¿ Zika rRT-PCR Test The LightMix¿ Zika rRT-PCR Test is a real-time RT-PCR test intended for the qualitative detection of RNA from the Zika virus in serum or EDTA plasma from individuals meeting CDC Zika virus clinical criteria (e.g., clinical signs and symptoms associated with Zika virus infection) and/or CDC Zika virus epidemiological criteria (e.g., history of residence in or travel to a geographic region with active Zika transmission at the time of travel, or other epidemiologic criteria for which Zika virus testing may be indicated). Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. ¿263a, to perform high complexity tests, or by similarly qualified non-U.S. laboratories. Results are for the identification of Zika virus RNA. Zika virus RNA is generally detectable in serum during the acute phase of infection (approximately 7 days following onset of symptoms, if present). Positive results are indicative of current infection. Laboratories are required to report all positive results to the appropriate public health authorities. Within the United States and its territories positive results must be reported to appropriate state/local public health authorities. Negative results do not preclude Zika virus infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information. The LightMix¿ Zika rRT-PCR Test is intended for use by trained clinical laboratory personnel specifically instructed and trained in the techniques of realtime PCR and in vitro diagnostic procedures. The Test is only for use under the Food and Drug Administration s Emergency Use Authorization.. Reason: The LightCycler¿ 480 algorithm used for the LightMix¿ Zika rRT-PCR Test, EUA (catalog number 07987897001) occasionally (~1% of the time) calls a positive result for a negative sample if there is an inflection point in the fluorescence signal..

Recalling firm
Roche Molecular Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-11-16
Terminated
2017-12-12
Firm location
Branchburg, NJ, United States

Product description

LightMix¿ Zika rRT-PCR Test The LightMix¿ Zika rRT-PCR Test is a real-time RT-PCR test intended for the qualitative detection of RNA from the Zika virus in serum or EDTA plasma from individuals meeting CDC Zika virus clinical criteria (e.g., clinical signs and symptoms associated with Zika virus infection) and/or CDC Zika virus epidemiological criteria (e.g., history of residence in or travel to a geographic region with active Zika transmission at the time of travel, or other epidemiologic criteria for which Zika virus testing may be indicated). Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. ¿263a, to perform high complexity tests, or by similarly qualified non-U.S. laboratories. Results are for the identification of Zika virus RNA. Zika virus RNA is generally detectable in serum during the acute phase of infection (approximately 7 days following onset of symptoms, if present). Positive results are indicative of current infection. Laboratories are required to report all positive results to the appropriate public health authorities. Within the United States and its territories positive results must be reported to appropriate state/local public health authorities. Negative results do not preclude Zika virus infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information. The LightMix¿ Zika rRT-PCR Test is intended for use by trained clinical laboratory personnel specifically instructed and trained in the techniques of realtime PCR and in vitro diagnostic procedures. The Test is only for use under the Food and Drug Administration s Emergency Use Authorization.

Reason for recall

The LightCycler¿ 480 algorithm used for the LightMix¿ Zika rRT-PCR Test, EUA (catalog number 07987897001) occasionally (~1% of the time) calls a positive result for a negative sample if there is an inflection point in the fluorescence signal.

Data sourced from openFDA. This site is unofficial and independent of the FDA.