Recall Z-0749-2024— Waldemar Link GmbH & Co. KG (Mfg Site)

Class II · Ongoing

Class II FDA medical device recall by Waldemar Link GmbH & Co. KG (Mfg Site), initiated 2023-11-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee arthroplasty procedure. Product Code: 15-2530/05. Reason: Mislabeled: Angle and height of the device marked as L(large) corresponds to the small/medium dimensions due to a product mix-up. This may result in too little bone material resected and not fit correctly leading to extended surgery time.

Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Classification
Class II
Status
Ongoing
Initiated
2023-11-30
Firm location
Norderstedt, N/A, Germany

Product description

Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee arthroplasty procedure. Product Code: 15-2530/05

Reason for recall

Mislabeled: Angle and height of the device marked as L(large) corresponds to the small/medium dimensions due to a product mix-up. This may result in too little bone material resected and not fit correctly leading to extended surgery time

Data sourced from openFDA. This site is unofficial and independent of the FDA.