Recall Z-0748-2021— Covidien Llc

Class II · Completed

Class II FDA medical device recall by Covidien Llc, initiated 2020-11-23. It names one FDA 510(k) clearance: K111825. Product: Covidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue. Item Number: TA30V3L. Reason: May contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis.

Recalling firm
Covidien Llc
Classification
Class II
Status
Completed
Initiated
2020-11-23
Firm location
Mansfield, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Covidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue. Item Number: TA30V3L

Reason for recall

May contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis