Recall Z-0747-2014— GE Healthcare, LLC
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2013-11-27, terminated 2016-09-07. It names one FDA 510(k) clearance: K071073. Product: GE Healthcare Carescape Patient Data Module The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes.. Reason: GE Healthcare has recently become aware of a potential safety issue due to intermittent failure of the spring loaded mounting latch of PDMs produced from April 2007 through December 2012..
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-11-27
- Terminated
- 2016-09-07
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Healthcare Carescape Patient Data Module The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes.
Reason for recall
GE Healthcare has recently become aware of a potential safety issue due to intermittent failure of the spring loaded mounting latch of PDMs produced from April 2007 through December 2012.