Recall Z-0745-2023— Heartware, Inc.

Class II · Ongoing

Class II FDA medical device recall by Heartware, Inc., initiated 2022-11-29. FDA's recall record names no specific 510(k) clearance for this event. Product: HeartWare HVAD Pump Kit, REF 1103. Reason: Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing..

Recalling firm
Heartware, Inc.
Classification
Class II
Status
Ongoing
Initiated
2022-11-29
Firm location
Miami Lakes, FL, United States

Product description

HeartWare HVAD Pump Kit, REF 1103

Reason for recall

Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.