Recall Z-0745-2020— Tosoh Smd Inc
Class III · Terminated
Class III FDA medical device recall by Tosoh Smd Inc, initiated 2019-11-15, terminated 2020-07-01. It names one FDA 510(k) clearance: K964348. Product: AIA-PACK B12 Pretreatment Set- IVD for diagnostic use for pre-treatment of calibrators/controls. Part Number: 020706. Reason: The AIA-2000 bar code scanners depend on the quiet zone to help identify the bar code. The lack of an adequate quiet zone can cause the bar code scanner to confuse the bar codes with text or graphics resulting in the scanner ignoring a valid symbol. This results in an error in reading the barcode: Error Code 2220 Unable to read barcode on reagent bottle. In some cases, the quiet zone is too short and an error in barcode reading occurs. The problem is intermittent and not all vials (6 vials are available per set) have been reported to fail. The only risk to patients or users is a delay in results. The company recommends (optional) that customers pre-screen the vials to identify affected products. Customers can contact the company for reimbursement of any affected products. The company has implemented barcode checks as a temporary solution and will replace the barcode printer on the production line for products beginning with lot B*01898..
- Recalling firm
- Tosoh Smd Inc
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2019-11-15
- Terminated
- 2020-07-01
- Firm location
- Grove City, OH, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
AIA-PACK B12 Pretreatment Set- IVD for diagnostic use for pre-treatment of calibrators/controls. Part Number: 020706
Reason for recall
The AIA-2000 bar code scanners depend on the quiet zone to help identify the bar code. The lack of an adequate quiet zone can cause the bar code scanner to confuse the bar codes with text or graphics resulting in the scanner ignoring a valid symbol. This results in an error in reading the barcode: Error Code 2220 Unable to read barcode on reagent bottle. In some cases, the quiet zone is too short and an error in barcode reading occurs. The problem is intermittent and not all vials (6 vials are available per set) have been reported to fail. The only risk to patients or users is a delay in results. The company recommends (optional) that customers pre-screen the vials to identify affected products. Customers can contact the company for reimbursement of any affected products. The company has implemented barcode checks as a temporary solution and will replace the barcode printer on the production line for products beginning with lot B*01898.
Data sourced from openFDA. This site is unofficial and independent of the FDA.