Recall Z-0745-2019— Bayer Medical Care, Inc.

Class II · Terminated

Class II FDA medical device recall by Bayer Medical Care, Inc., initiated 2018-12-14, terminated 2020-06-16. It names 4 FDA 510(k) clearances: K173773, K112086, K023183 and 1 more. Product: MAVIG Portegra or Portegra2 Overhead Counterpoise System (OCS) used in conjunction with the Medrad Stellant CT Injection System (Certegra Workstation), Medrad Mark 7 Arterion" Injection System, or Medrad Mark V ProVis Injection System, Catalog Numbers: Catalog numbers: OCS CEIL (58-S, 58-P, 85-S, 85-P, 100-S, 100-P, 70Si, 97Si), OCS TRACK (58-S, 58-P, 80-S, 80-P), OCS WALL (S, P), OCS (115A, 115CT, 125CT, 125A, 135A, 135CT, 145, 155A, 165A, 115AS, 115CTS), EOC (700, 700S), EOW 700, EOA 700. Reason: If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall..

Recalling firm
Bayer Medical Care, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-12-14
Terminated
2020-06-16
Firm location
Indianola, PA, United States

510(k) clearances named in this recall

4 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

MAVIG Portegra or Portegra2 Overhead Counterpoise System (OCS) used in conjunction with the Medrad Stellant CT Injection System (Certegra Workstation), Medrad Mark 7 Arterion" Injection System, or Medrad Mark V ProVis Injection System, Catalog Numbers: Catalog numbers: OCS CEIL (58-S, 58-P, 85-S, 85-P, 100-S, 100-P, 70Si, 97Si), OCS TRACK (58-S, 58-P, 80-S, 80-P), OCS WALL (S, P), OCS (115A, 115CT, 125CT, 125A, 135A, 135CT, 145, 155A, 165A, 115AS, 115CTS), EOC (700, 700S), EOW 700, EOA 700

Reason for recall

If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.