Recall Z-0741-2019— Draeger Medical, Inc.

Class I · Terminated

Class I FDA medical device recall by Draeger Medical, Inc., initiated 2018-12-21, terminated 2019-04-16. FDA's recall record names no specific 510(k) clearance for this event. Product: VentStar Anesthesia WT(P) 180, Catalog Number MP00374 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of condensate. The standard breathing circuit is equipped with a water trap in each breathing hose limb.. Reason: Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit.

Recalling firm
Draeger Medical, Inc.
Classification
Class I
Status
Terminated
Initiated
2018-12-21
Terminated
2019-04-16
Firm location
Telford, PA, United States

Product description

VentStar Anesthesia WT(P) 180, Catalog Number MP00374 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of condensate. The standard breathing circuit is equipped with a water trap in each breathing hose limb.

Reason for recall

Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit

Data sourced from openFDA. This site is unofficial and independent of the FDA.