Recall Z-0735-2022— Ortho-Clinical Diagnostics, Inc.
Class II · Ongoing
Class II FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2021-12-10. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.. Reason: The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value..
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-12-10
- Firm location
- Rochester, NY, United States
Product description
VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.
Reason for recall
The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value.