Recall Z-0734-2023— Fresenius Medical Care Holdings, Inc.

Class II · Ongoing

Class II FDA medical device recall by Fresenius Medical Care Holdings, Inc., initiated 2022-07-27. It names one FDA 510(k) clearance: K173972. Product: 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR. Reason: (1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device"..

Recalling firm
Fresenius Medical Care Holdings, Inc.
Classification
Class II
Status
Ongoing
Initiated
2022-07-27
Firm location
Waltham, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR

Reason for recall

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

Data sourced from openFDA. This site is unofficial and independent of the FDA.