Recall Z-0729-2023— GE Medical Systems, LLC
Class II · Ongoing
Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2022-11-16. It names one FDA 510(k) clearance: K213490. Product: B105P Patient Monitor (with E-module slot option), REF 6160000-001-XXXXXX,. Reason: GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level..
- Recalling firm
- GE Medical Systems, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-11-16
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
B105P Patient Monitor (with E-module slot option), REF 6160000-001-XXXXXX,
Reason for recall
GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level.
Data sourced from openFDA. This site is unofficial and independent of the FDA.