Recall Z-0726-2024— Limacorporate S.p.A
Class II · Ongoing
Class II FDA medical device recall by Limacorporate S.p.A, initiated 2023-08-29. It names one FDA 510(k) clearance: K222427. Product: PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870. Reason: Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position..
- Recalling firm
- Limacorporate S.p.A
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-08-29
- Firm location
- San Daniele Del Friuli, N/A, Italy
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870
Reason for recall
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.