Recall Z-0723-2025— MEDLINE INDUSTRIES, LP - Northfield

Class II · Ongoing

Class II FDA medical device recall by MEDLINE INDUSTRIES, LP - Northfield, initiated 2024-11-06. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline brand, medical procedure convenience kits, labeled as: 1) CRANIOTOMY, SKU CDS982719X; 2) MVFF/FIBU #54-RF, SKU DYNJ21927Q; 3) MVFF/FIBU #54-RF, SKU DYNJ21927S; 3) NEURO PEDIATRIC, SKU DYNJ61607A; 4) CRANIOTOMY PACK, SKU DYNJ66419C; 5) LYMPHATIC BYPASS PACK, SKU DYNJ69907; 6) LYMPHATIC BYPASS PACK, SKU DYNJ69907A; 7) HEAD AND NECK PACK - 51597, SKU DYNJ81934B; 8) MAJOR CRANIOTOMY PACK, SKU DYNJ82007A; 9) SPINE PACK, SKU DYNJ82011A; 10) CRANIOTOMY MERCY STK 256, SKU DYNJ84068; 11) WMC UNIVERSAL NEURO PACK, SKU DYNJ84496A; 12) SPINE PACK LF, SKU DYNJ85830; 13) AHT SPINE PACK, SKU DYNJ903566D; 14) LAMINECTOMY SPINAL FUSION PACK, SKU DYNJ909364A; 15) DIEP, SKU DYNJ909472; 16) AMBULATORY SINUS, SKU DYNJ910292; 17) AMBULATORY SINUS, SKU DYNJ910292A. Reason: Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods.

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Classification
Class II
Status
Ongoing
Initiated
2024-11-06
Firm location
Northfield, IL, United States

Product description

Medline brand, medical procedure convenience kits, labeled as: 1) CRANIOTOMY, SKU CDS982719X; 2) MVFF/FIBU #54-RF, SKU DYNJ21927Q; 3) MVFF/FIBU #54-RF, SKU DYNJ21927S; 3) NEURO PEDIATRIC, SKU DYNJ61607A; 4) CRANIOTOMY PACK, SKU DYNJ66419C; 5) LYMPHATIC BYPASS PACK, SKU DYNJ69907; 6) LYMPHATIC BYPASS PACK, SKU DYNJ69907A; 7) HEAD AND NECK PACK - 51597, SKU DYNJ81934B; 8) MAJOR CRANIOTOMY PACK, SKU DYNJ82007A; 9) SPINE PACK, SKU DYNJ82011A; 10) CRANIOTOMY MERCY STK 256, SKU DYNJ84068; 11) WMC UNIVERSAL NEURO PACK, SKU DYNJ84496A; 12) SPINE PACK LF, SKU DYNJ85830; 13) AHT SPINE PACK, SKU DYNJ903566D; 14) LAMINECTOMY SPINAL FUSION PACK, SKU DYNJ909364A; 15) DIEP, SKU DYNJ909472; 16) AMBULATORY SINUS, SKU DYNJ910292; 17) AMBULATORY SINUS, SKU DYNJ910292A

Reason for recall

Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods

Data sourced from openFDA. This site is unofficial and independent of the FDA.