Recall Z-0721-2017— RAYSEARCH LABORATORIES AB
Class II · Terminated
Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2016-11-09, terminated 2019-06-11. It names one FDA 510(k) clearance: K160093. Product: RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.. Reason: Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected..
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-11-09
- Terminated
- 2019-06-11
- Firm location
- Stockholm, N/A, Sweden
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Reason for recall
Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected.