Recall Z-0719-2021— B Braun Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by B Braun Medical, Inc., initiated 2020-11-23, terminated 2023-03-14. FDA's recall record names no specific 510(k) clearance for this event. Product: APEX Compounding System Control Panel Module, REF: 601238, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.. Reason: There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a a compounded drug that does not meet the specified accuracy range for an individual ingredient..

Recalling firm
B Braun Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-11-23
Terminated
2023-03-14
Firm location
Carrollton, TX, United States

Product description

APEX Compounding System Control Panel Module, REF: 601238, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.

Reason for recall

There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a a compounded drug that does not meet the specified accuracy range for an individual ingredient.

Data sourced from openFDA. This site is unofficial and independent of the FDA.