Recall Z-0715-2024— Philips North America

Class II · Ongoing

Class II FDA medical device recall by Philips North America, initiated 2021-07-15. It names one FDA 510(k) clearance: K213583. Product: (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359). Reason: The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.).

Recalling firm
Philips North America
Classification
Class II
Status
Ongoing
Initiated
2021-07-15
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

(1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)

Reason for recall

The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)