Recall Z-0713-2026— B Braun Medical Inc
Class II · Ongoing
Class II FDA medical device recall by B Braun Medical Inc, initiated 2025-10-29. It names one FDA 510(k) clearance: K931377. Product: SAFELINE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. UNIVERSAL IV ADMIN SET; Catalog Number: 352604. 2. SAFELINE ADDITIV ALTVTD 3 INJ S; Primary Gravity IV Set with 3 SafeLine Injection Sites; Catalog Number: NF1251. 3. SAFELINE PED. ADDITIVE; Primary Gravity IV Set with 3 SafeLine Injection Sites; Catalog Number: NF1290.. Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion)..
- Recalling firm
- B Braun Medical Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-10-29
- Firm location
- Bethlehem, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
SAFELINE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. UNIVERSAL IV ADMIN SET; Catalog Number: 352604. 2. SAFELINE ADDITIV ALTVTD 3 INJ S; Primary Gravity IV Set with 3 SafeLine Injection Sites; Catalog Number: NF1251. 3. SAFELINE PED. ADDITIVE; Primary Gravity IV Set with 3 SafeLine Injection Sites; Catalog Number: NF1290.
Reason for recall
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).