Recall Z-0713-2014— Radiometer America Inc

Class II · Terminated

Class II FDA medical device recall by Radiometer America Inc, initiated 2013-06-10, terminated 2014-06-19. It names 3 FDA 510(k) clearances: K935655, K922922, K895181. Product: pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).. Reason: RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration..

Recalling firm
Radiometer America Inc
Classification
Class II
Status
Terminated
Initiated
2013-06-10
Terminated
2014-06-19
Firm location
Westlake, OH, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).

Reason for recall

RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.